6 Remote Operations
62 Keysight J7201A/B/C Operating and Service Manual
Examples
To query the corrected value:
CORRection? 25dB,6GHz
Related Commands
CORRection:AUTO?
[INPut:]ATTenuation:AUTOcorrected/INPut[:ATTenuation]:AUTOcorrected
CORRection:AUTO?
Reads the corrected value in auto mode.
CORRection:AUTO?
Examples
To query the corrected value:
CORRection:AUTO?
Related Commands
CORRection?
[INPut:]ATTenuation:AUTOcorrected/INPut[:ATTenuation]:AUTOcorrected
DISPlay:ATTenuation
Turns the front panel display to the ATTEN page.
DISPlay:ATTenuation
Maximum
121dB (J7201A/B) – attenuation level
101dB (J7201C) – attenuation level
6 GHz (J7201A) – for frequency
18 GHz (J7201B) – for frequency
26.5 GHz (J7201C) – for frequency
Units dB, Hz
Query? Yes, query only
Set? No
Return Parameters (Optional)
Numeric representation format: This
parameter is a number that can be expressed
as an integer (e.g., 8) or a real number (e.g.,
23.6)
Minimum n/a
Maximum n/a
Units None
Query? No
Set? Yes