Remote Operations 6
Keysight J7201A/B/C Operating and Service Manual 67
Examples
To query the next error from the queue:
SYST:ERR?
Related Commands
SYSTem:PREset
SYSTem:REMote
SYSTem:COMMunicate:TCPip:CONTrol?
SYSTem:PREset
Presets the attenuation control unit to the following factory-defined states:
- :ATT 121 (For J7201A/B)
- :ATT 101 (For J7201C)
SYSTem:PREset
Examples
To preset the attenuation control unit:
SYSTem:PREset
Related Commands
SYSTem:ERRor?
SYSTem:REMote
SYSTem:COMMunicate:TCPip:CONTrol?
SYSTem:REMote
Sets the interface in the Remote state, which disables all front panel controls except the
Local key. Pressing the Local key while in the Remote state returns the front panel to the
Local state.
SYSTem:REMote
Query? No
Set? Yes
Set Parameters n/a
Minimum n/a
Maximum n/a
Units None
Query? No
Set? Yes
Set Parameters n/a
Minimum n/a