6 Remote Operations
68 Keysight J7201A/B/C Operating and Service Manual
Examples
To set the interface in the Remote state:
SYST:REM
Related Commands
SYSTem:ERRor?
SYSTem:PREset
SYSTem:COMMunicate:TCPip:CONTrol?
SYSTem:COMMunicate:TCPip:CONTrol?
Reads the initial control connection port number for Sockets communications. This
connection is used to send and receive commands and queries.
- Use the Control socket connection to send a Device Clear (DCL) to the instrument or to
detect pending Service Request (SRQ) events.
SYSTem:COMMunicate:TCPip:CONTrol?
Examples
To query the initial control connection port number for Sockets communications:
SYST:COMM:TCP:CONT?
Related Commands
SYSTem:ERRor?
SYSTem:PREset
SYSTem:REMote
SYSTem:PERSona:MANufacturer
Sets the manufacturer string returned by *IDN? Query.
SYSTem:PERSona:MANufacturer <quoted string>
Maximum n/a
Units None
Query? Yes, query only.
Set? No
Return Parameters Integer
Minimum n/a
Maximum n/a
Units None
Query? Yes
Set? Yes
Set/Return Parameters String