EasyManuals Logo

Keysight Technologies N5172B EXG Service Guide

Keysight Technologies N5172B EXG
451 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #420 background imageLoading...
Page #420 background image
420 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Pre and Post-Repair Procedures
Additional Tasks
Figure 16-1 Factory I/Q Calibration Performed
4. If there is no Date, Range, and Method on the display that the Factory
Calibration was performed, as shown in Figure 16-2, the Enhanced Factory
Calibration will need to be run.
Figure 16-2 Factory I/Q Calibration Not Performed
5. Press Enhanced Factory Calibration.
6. Connect and configure a supported power meter and spectrum analyzer to
the signal source, or a communication bus that it supports. See Chapter 5,
“Service and Utility Menus,” for detailed information on configuring this
equipment.
7. Once the required equipment has been connected and verified press
Execute Enhanced Int Chan Correction Cal.
For detailed information on running the Enhanced Factory Calibration routine
see Chapter 5, “Service and Utility Menus”
This procedure could take up to 1¼ hours or more to complete. Once this
procedure has completed continue with the next task associated with the
assembly being replaced.

Table of Contents

Other manuals for Keysight Technologies N5172B EXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5172B EXG and is the answer not in the manual?

Keysight Technologies N5172B EXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5172B EXG
CategoryPortable Generator
LanguageEnglish

Related product manuals