Keysight Agile Signal Generator Service Guide 101
Assembly Replacement
Inner Bottom Covers
Tools Required
T-10 driver
Removal Procedure
Refer to Figure 3-4 and Figure 3-5 for this procedure.
1. Disconnect the power cord.
2. Remove the outer-cover from the signal generator. Refer to “Outer Cover”
on page 99.
3. Place the signal generator upside down.
4. Refer to Figure 3-4. To remove the rear bottom cover (1), use the T-10
driver to remove the twelve screws (2) (0515-1946) (including two on the
rear panel).
5. Refer to Figure 3-5 To remove the front bottom cover (3), use the T-10
driver to remove the sixteen screws (4) (0515-1946).
Replacement Procedure
1. Reverse the order of the removal procedures.
2. Recommended torque for all screws is 9 in–lbs.