Keysight Agile Signal Generator Service Guide 153
Assembly Replacement
Replacement Procedure
1. Reverse the order of the removal procedures.
2. Refer to Figure 3-48. Torque screw (3) (0515-0430) to 6 in-lbs.
Recommended torque for all other screws is 9 inlbs.
3. Refer to Figure 3-47. Using the 5/8-inch nut driver, torque the connector
nuts (1) to 21 in-lbs. Using the torque open-end wrench, torque W9 cable
(2) to 9 in-lbs. Using the hex driver torque the hex nuts (3) to 9 in-lbs.
4. Refer to Figure 3-46. When replacing the A8 Solid State Drive (1) and the
A13 Module (2), torque the thumb screw to 9 in-lbs.