Keysight Agile Signal Generator Service Guide 103
Assembly Replacement
Front Frame
Tools Required
T-20 driver
needle-nose pliers
socket wrench and 1/2-inch slotted deep socket
Removal Procedure
1. Disconnect the power cord.
2. Remove the outer-cover from the signal generator. Refer to “Outer Cover”
on page 99.
3. Place the signal generator upside down on a flat surface.
4. Remove the rear bottom cover from the signal generator. Refer to “Inner
Bottom Covers” on page 101.
5. Refer to Figure 3-6. Using the T-10 driver, remove the eleven screws (1)
(0515-1946) that attach the left side cover to the chassis. Remove the left
side cover.
Figure 3-6 Left Side Cover Removal
Use a soft surface or a soft cloth to protect the front frame while performing this procedure.