154 Keysight Agile Signal Generator Service Guide
Assembly Replacement
RF Connector
Tools Required
T-20 driver
5/16-in open-end wrench driver
Removal Procedure
1. Disconnect the power cord.
2. Remove the outer-cover from the signal generator. Refer to “Outer Cover”
on page 99.
3. Perform step 1 through step 9 of the front frame removal procedure. It is
not necessary to remove W33. Refer to “Front Frame” on page 103.
4. Refer to Figure 3-49. Loosen the nut on the semi-rigid cable connected to
the RF connector.
Figure 3-49 Loosen Semi-Rigid Cable