Front Panel Menu Reference SCPI Command
Not available To query how many (if any) additional triggers occurred:
TRIG:ACQ:IND:COUN?
To return the indices where the triggers occurred:
TRIG:ACQ:IND?
You can also return the actual measurement data that was captured after any of the aforementioned
trigger indices.
Front Panel Menu Reference SCPI Command
Not available To return DC voltage or current calculated after the trigger indcies:
FETC:VOLT? [<start_index>, <points>]
FETC:CURR? [<start_index>, <points>]
To return instantaneous voltage or current data after the trigger
indices:
FETC:ARR:VOLT? [<start_index>, <points>]
FETC:ARR:CURR? [<start_index>, <points>]
4 Using the Regenerative Power System
154 Keysight RP7900 Series Operating and Service Guide