EasyManua.ls Logo

Keysight 3458A - Inguard Power Supplies Troubleshooting

Keysight 3458A
137 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Assembly Level Troubleshooting 4
Keysight 3458A Assembly-Level Repair Manual 135
7 If the power supply voltages are now good, the Outguard Controller (A5)
assembly is defective. Go to the Assemblies Removal/Installation Procedures
in Chapter 3 of this manual to replace the assembly.
8 If the voltage is still low, the Outguard Power Supply (A6) assembly is
defective. Go to the Assemblies Removal/Installation Procedures in Chapter 3
of this manual to replace the assembly.
Inguard power supplies troubleshooting
Do the following:
1 Use the Covers Removal/Installation Procedures in Chapter 3 of this manual to
remove the instrument's bottom cover and bottom shield, if not previously
removed.
2 Refer to Figure 4-3. With the test digital multimeter, check for
+5 V (±0.13 V), +18 V (±1.8 V), and - 18 V (±1.8 V) at the test points shown in
the figure.
3 If all voltages are good, the power supply assembly is good.
4 If a voltage is high, the Inguard Power Supply (A4) assembly is defective. Go to
the Assemblies Removal/Installation Procedures in Chapter 3 of this manual to
replace the assembly.
5 If a voltage is low, other assemblies in the inguard may be loading down the
power supply. Continue with the next step.
6 Refer to Figure 4-3 to locate the gray 20-pin cable that connects the Inguard
Power Supply (A4) assembly to the DC Circuitry (A1) assembly. Unplug the
cable at the power supply assembly.
7 If the power supply voltage that was low is now good, the DC Circuitry (A1)
assembly is defective. Go to the Assemblies Removal/Installation Procedures
in Chapter 3 of this manual to replace the assembly.

Table of Contents

Other manuals for Keysight 3458A

Related product manuals