4 Assembly Level Troubleshooting
112 Keysight 3458A Assembly-Level Repair Manual
3458A Adjustments/Calibration
Instrument adjustments/calibration must be performed whenever a printed circuit
board assembly has been replaced or removed from the instrument. Removal or
replacement of some assemblies only requires internal calibration (by executing
the “ACAL ALL” command). Others need complete adjustments (for example,
Offset, DC Gain, etc.). It is also good practice to execute the “ACAL ALL”
command after the covers have been removed and replaced on the instrument.
Table 4-1 lists the 3458A assemblies and the required adjustments/calibration
after assembly replacement. For information on how to adjust/calibrate the
3458A, refer to the 3458A Calibration Manual.
Table 4-1 Adjustments/Calibration needed after assembly replacement
Ref
desig.
Keysight part
number
Assembly description Adjustments needed
A1 03458-66501 DC circuitry
Offset, DC Gain, Resistance, and DC
Current Adjustments
A2 03458-66502 AC converter AC Adjustment
A3 03458-66503 A/D converter and inguard logic
Offset, DC Gain, Resistance, DC
Current, and AC Adjustments
A4 03458-66504 Inguard power supply ACAL ALL
A5 03458-66505 Outguard controller
Offset, DC Gain, Resistance, DC
Current, and AC Adjustments
A5 03458-66515 Outguard controller (Opt 001)
Offset, DC Gain, Resistance, DC
Current, and AC Adjustments
A6 03458-66506 Outguard power supply ACAL ALL
A7 03458-66507 Display logic ACAL ALL
A9 03458-66509 DC reference
Offset, DC Gain, Resistance, and DC
Current Adjustments
A9 03458-66519 DC reference (Opt 002)
Offset, DC Gain, Resistance, and DC
Current Adjustments
A10 03458-66510 Front/rear switch Offset Adjustments, ACAL ALL