100 Keysight CX3300 User’s Guide, Edition 4
Performing User Calibration
Using the CX1151A
Figure 3-7 Connection Diagram for User Calibration (CX1151A)
After the channel passed the user calibration, the ON/OFF switch will be
available.
Set ON to enable error correction.
Error correction is performed for the data measured by this channel from now
on.
Set OFF to disable error correction.
Error correction is not performed.
If the channel failed th
e user calibration, the ON/OFF switch is not available.
Error correction is not performed.
Once you disconnect the CX1151A from the CX3300, the user calibration data
for this channel will be cleared.