Keysight CX3300 User’s Guide, Edition 4 95
Performing User Calibration
Using the CX1104A
Using the CX1104A
The CX3300 provides the offset user calibration for the CX1104A.
The user calibration data, Offset [A], is acquired and is used for error correction.
Effective for reducing the offset error.
Preparation: Before starting the user calibration, connect the CX1104A to the CX3300 analog
input channel.
Prepare a resistive sensor head (CX1211A,
CX1212A, CX1213A, CX1214A,
CX1215A, or CX1216A).
After the connection, warm up them for 30 minutes.
Procedure: You can perform the user calibration as shown below.
1. Press the
Menu key several times to open the Configuration dialog box. The
Menu key is located above the Horizontal control on the front panel.
2. Click User Calibration to display the Configuration > User Calibration screen.
3. Click the Ex
ecute button for the channel under user calibration.
You will see a dialog box which asks you to connect a resistive sensor head to
the CX1104A input using the extension cable. Then connect them as shown
in Figure 3-5 and click Yes.
The user calibration will start.