Keysight CX3300 Device Current Waveform Analyzer
User’s Guide
3
Performing User Calibration
Using the CX1101A/CX1102A/CX1103A 86
Using the CX1104A 95
Using the CX1105A 97
Using the CX1151A 99
User calibration allows you to perform error correction of measurement data.
This is effective for making more accurate measurements. This chapter describes
how to perform the user calibration.
The user calibration must be performed
after warming-up of 30 minutes.
The user calibration is effective for:
the environmental temperature of 0 to 40 C
3 C from the temperature when the last user calibration is performed
24 hours since the last user calibration is performed
All of the user calibration data is cleared by applying the factory default. The
data is not cleared by turning the CX3300 off.