Keysight CX3300 User’s Guide, Edition 4 23
Setting Up the CX3300
Verifying Basic Operation
Verifying Basic Operation
The following procedures verify the basic operation which monitors the
Calibration Output signal.
• “If You Use the CX1101A/CX1102A/CX1103A”
• “If You Use the CX1104A”
• “If You Use the CX1105A”
If the CX1100 has not been connected yet, connect it. See “Connecting the
CX1100” on page 21.
If You Use the CX1101A/CX1102A/CX1103A
Required accessories
• CX1203A sensor head, 1 ea., furnished with the CX1101A/CX1102A, no need
for the CX1103A
• Adapter (SMA plug to BNC jack), 1 ea., furnished with the current sensor
• Coaxial cable (BNC plug to BNC plug), 1 ea., furnished with the CX3300
Procedure
1. Connect the current sensor to the Aux Out terminal on the CX3300 front
panel by using the accessories listed above.
2. Press the
1 (yellow), 2 (green), 3 (blue), or 4 (red) key associated with the
channel connected to the Aux Out terminal and enable the channel.
3. Press the
Menu key several times to open the Configuration dialog box. The
Menu key is located above the Horizontal control on the front panel.
4. Set the following conditions on the Configuration dialog box (Configuration >
Calibration Output).
• Output State: ON
• Source Shape: DC
• Source Mode: Cu
rrent
• Current: 2 mA