Keysight E4980A/E4980AL Precision LCR Meter 51
Troubleshooting
Function Specific Troubleshooting
Figure 4-7 Network Dialog Box (“IP Address” Tab)
Step 8. By clicking “OK” button, the network settings take effect. Verify the E4980A/
E4980AL LAN settings and response to a command from the external
computer as follows:
a. On the external computer, click "Start" button in the Windows Start menu
bar using a mouse.
b. Click "Programs" and "MS-DOS Prompt" (or "Command Prompt") to open
the DOS (Command) Prompt window.
c. Type a ping command followed by the E4980A/ E4980AL IP address as
"ping xxx.xxx.xxx.xxx" (where, xxx.xxx.xxx.xxx is the IP address such as
192.168.0.1 assigned in step 3-d.) Press [Enter] key on the keyboard.
d. The external computer displays a message of ping command response
from the E4980A/ E4980AL.
Figure 4-8 shows an example of normal result
that indicates the IP address, packet size, and a response time counted by
the computer. The ping command is repeated 4 times.
Figure 4-8 Example of normal ping command test result