Testing Performance 3
S-Series Oscilloscopes Service Guide 35
Connections
NOTE
Connect output 1 of the 11667B splitter to the oscilloscope Channel n input directly using the
1250-1200 adapter, without any additional cabling or adapters.
Connect output 2 of the 11667B power splitter to the power sensor using the 1250-1750 N-type to
3.5mm adapter, without any additional cabling or adapters.
Minimize the use of other adapters.
Ensure that SMA and 3.5 mm connectors are tightened properly:
8 in-lbs (90 N-cm) for 3.5 mm
5 in-lbs (56 N-cm) for SMA