EasyManuals Logo

Keysight Infiniium MSOS404A Service Guide

Keysight Infiniium MSOS404A
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #37 background imageLoading...
Page #37 background image
Testing Performance 3
S-Series Oscilloscopes Service Guide 37
T
8 Measure the input power to the oscilloscope channel and convert this
measurement to Volts RMS using the expression:
For example, if the power meter reading is 4.0 μW, then Vin = (4.0*10
-6
*
50Ω)
1/2
= 14.1 mVrms.
Record the RMS voltage in the Analog Bandwidth—Maximum Frequency Check
section of the Performance Test Record (in the Vin @ 50 MHz column).
9 Press [Clear Display] on the oscilloscope and record the scope V rms reading in
the Analog Bandwidth—Maximum Frequency Test section of the Performance
Test Record (Vout @ 50 MHz).
Nominal Generator Amplitude Settings
Oscilloscope
Vertical Scale
Generator Signal
Amplitude (Vp-p)
Generator Signal
Amplitude (dBm)
5 mV/div 0.02 -30
10 mV/div 0.04 -24
20 mV/div 0.08 -18
50 mV/div 0.20 -10
100 mV/div 0.40 -4
200 mV/div 0.80 +2
500 mV/div 2.0 +10
1 V/div 4.0 +16
V
in
P
meas
50Ω×=
NOTE
For all oscilloscope readings in this procedure, use the mean value in the Measurement Results area
at the bottom of the screen.
If a question mark appears in front of any value at the bottom of the screen, press [Clear Display] on
the oscilloscope, wait for the number of averages to return to 16, and then record the oscilloscope
reading.

Table of Contents

Other manuals for Keysight Infiniium MSOS404A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Infiniium MSOS404A and is the answer not in the manual?

Keysight Infiniium MSOS404A Specifications

General IconGeneral
BrandKeysight
ModelInfiniium MSOS404A
CategoryTest Equipment
LanguageEnglish

Related product manuals