B - Performance Test and Calibration
90
N330xA Module (top view)
Figure B-1. I MON ZERO CALIBRATION
Keysight 3458 or
34401 DVM
INPUT TERMINALS
MUST BE OPEN
+
_
N330xA Module (rear view)
Figure B-2. I MON / EXT PRG and CURRENT
CALIBRATION
Keysight 3458 or
34401 DVM
A11
A9
Iprog Power Source +
& DVM -
A8
A7
Keysight 3458 or
34401 DVM
If 3 DVM's are not available - move DVM
between I MON, EXT PRG and current shunt as
required by program
SHUNT
Power
Source
+
_
I MON
A COM
EXT PRG
EXT COM
I MON
A COM
-
+
GND_AQS
IMON_INV
Location of I MON zero pins
on single-slot modules
IMON_INV
GND_AQS
Ribbon
Cable
Note: Use this equipment
only when calibrating
Imon and Iprog gain.