Performance Test and Calibration - B
91
+
_
N330xA Module
Figure B-3. VOLTAGE CALIBRATION
Keysight 3458 or
34401 DVM
+
_
N330xA Module
Figure B-4. RESISTANCE CALIBRATION
Keysight 3458 or
34401 DVM
+S
-S
Keysight 3458 or
34401 DVM
SHUNT
+S
-S
If 2 DVM's are not available - move DVM between
+/- sense and shunt as required by program
Be sure sense switch is in RMT position
Be sure sense switch is in RMT position