EasyManua.ls Logo

Keysight N8976B

Keysight N8976B
524 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight NFA Series Noise Figure Analyzers Service Guide 163
RF Section Troubleshooting (N8973B, 74B, 75B Analyzers)
Troubleshooting
The level should be 25 dBm ± 2 dB as shown in Figure 4-14.
Figure 4-14 50 MHz Calibrator Signal on Output of Attenuator A
Press Mech Atten and enter 2 dB. The 50 MHz calibrator signal measured on
the functioning Spectrum Analyzer should measure 2 dB lower than the
previous step (~27 dBm). Press Mech Atten and enter 4 dB. The 50 MHz
calibrator signal measured on the functioning Spectrum Analyzer should
measure an additional 2 dB lower than the previous step (~29 dBm). If the
power levels measure correctly, reconnect W11 cable. If either of these levels is
incorrect, Input Attenuator A is the most probable cause, provided the control
logic from the A15 Front End Control Assembly was previously verified.

Table of Contents

Other manuals for Keysight N8976B

Related product manuals