Keysight NFA Series Noise Figure Analyzers Service Guide 223
RF Section Troubleshooting (N8976B)
Troubleshooting
Attenuation Check
Set the signal source connected to the analyzer input port to 0 dBm. On the
analyzer, press Amplitude, Attenuation, Mech Atten and change input
attenuation to 0 dB. The measuring spectrum analyzer connected to A10
Attenuator output port should indicate a 0 dBm level. Change the input
attenuation on the analyzer under test to 2 dB.
See the chart below for expected measurement values at the A10 Attenuator
output port.
Analyzer Under Test
Attenuator Setting (dB)
Power at A10 Output
Port (dBm)
Input Attenuator Being
Tested
0 0 dBm (reference) Both set to through path
2 2A9
4 4A9
6 −6 A9
8 8A9
10 10 A10
20 20 A10
30 30 A10
40 40 A10
50 50 A10
60 60 A10