Keysight N9010B EXA Signal Analyzer Service Guide   103
Instrument Messages
Event Messages
–233 Invalid version A legal data element was found but could not be used because the version of 
the data is incorrect. For example, state data changes as new instrument 
features are added, so old state files may not work in an instrument with a 
newer version of software
–240 Hardware error A legal program command or query could not be executed because of a 
hardware error. The exact problem cannot be specifically identified
–240 Hardware error; See 
details in Windows 
Event Log under SA
The internal data acquisition system detected a problem at startup and logged 
the details in the Windows Event Log
–241 Hardware missing The operation could not be performed because of missing hardware; perhaps 
the optional hardware is not installed
–241 Hardware missing; 
Input not available
The hardware required is not part of this model or the option is not installed
–241 Hardware missing; 
Internal preamp not 
available at all 
frequency points
The Internal Preamp is currently turned on, but the measurement is being 
performed completely or partially outside the range of the preamp. It is 
recommended that the user turns preamp off to ensure consistent results 
across the entire measurement
–241 Hardware missing; not 
available for this 
model number
The hardware required is not part of this model
–241 Hardware missing; 
Option not installed
The optional hardware is not installed
–250 Mass storage error; A problem was found with the mass storage device (memory, disk drive, etc.). 
The exact problem cannot be specifically identified
–250 Mass storage error; 
Access denied
Access is denied
–250 Mass storage error; 
Bad path name
The specified path is invalid
–250 Mass storage error; 
Can only import 
single trace .csv 
files
Trace files containing multiple traces can not be imported. However, if you 
need to recall multiple traces you can use the 
Save and Recall functions, 
rather than the Import and Export functions
–250 Mass storage error; 
Can only load an 
Antenna Unit into 
Correction 1
The only Correction register that supports Antenna Units is number 1. You have 
attempted to load an Ampcor file that contains antenna units into another 
register
–250 Mass storage error; 
Cannot make
The directory or file cannot be created
–250 Mass storage error; 
Different Antenna 
Unit already in use
Attempt to import Corrections file with Antenna Unit that differs from an in-use 
correction
Err# Message Verbose/Correction Information