102 Keysight N9010B EXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–224 Illegal parameter 
value; Measurement 
not available
You tried to turn on a measurement that is not available in the current mode
–224 Illegal parameter 
value;This instrument 
is always DC coupled
You cannot set AC coupling in this instrument
–224 Illegal parameter 
value;This model is 
always AC coupled
You cannot set DC coupling in this analyzer
–225 Out of memory There is not enough memory to perform the requested operation
–225 Out of memory; 
Insufficient 
resources to load 
Mode <mode name>
If you attempt to load a mode via SCPI that will exceed memory capacity, the 
Mode does not load and this message is returned. “mode name” is the SCPI 
parameter for the Mode in question (for example, SA for Spectrum Analyzer 
Mode.) You can free up resources using the 
System, Power On, 
Configure Applications
 menu
–225 Out of memory; Memory 
limit caused Data 
Acquisition to be 
truncated
–226 List not same length You are using the LIST structure, but have individual lists of differing lengths
–230 Data corrupt or 
stale;
A legal data element was found, but it could not be used because the data 
format or the data structure was not correct. This could occur if a new 
measurement had been started but had not completed
–230 Data corrupt or 
stale; Measurement 
data is not available
Measurement data not available. The measurement that you are trying to get 
data from must be the current active measurement. Maybe you have not 
initiated the measurement, or it has not completed all the sweeps/averages 
needed
–230 Data corrupt or 
stale; Trace contains 
no data.
Trace cannot be displayed because currently there is no data assigned to it. 
Use the functions under the 
Trace menu, or load a previously saved trace, to 
assign data to the trace
–230 Data corrupt or 
stale; Unable to load 
state from file
There is something wrong with the state data in the desired file. Maybe the file 
is corrupt, or it is from an instrument/version that is not recognized by the 
current instrument
–231 Data questionable Indicates that the measurement accuracy is suspect
–232 Invalid format A data element was found but it could not be used because the data format or 
the data structure was not correct
–232 Invalid format; Map 
information not 
loaded
Instrument failed to load the burst mapping information from the selected file