Keysight N9010B EXA Signal Analyzer Service Guide 101
Instrument Messages
Event Messages
–224 Illegal parameter
value; Exceeding the
max list length
The list parameters have a maximum allowed length. You are trying to set a
length longer than the maximum
–224 Illegal parameter
value; Gated FFT is
not available while
Sweep Type is set to
Swept
The gated FFT function is not available if you have selected the swept type of
sweep. You must select the FFT sweep type
–224 Illegal parameter
value; Gated LO is
not available while
Sweep Type is set to
FFT
The FFT sweep type moves the LO frequency in steps, so the gated LO function
is not available if you have selected FFT sweep
–224 Illegal parameter
value; Gated Video is
not available while
Sweep Type is set to
FFT
The gated video function is not available if you have selected the FFT sweep
type
–224 Illegal parameter
value; Illegal
identifier
<identifier>. This
value may already be
in use.
The value for the LXI LAN identifier parameter must be unique (that is, LAN0
and LAN7 must have different identifier strings)
–224 Illegal parameter
value; Index out of
range
When querying the LXI Event Log or the Servo Log, an index may be used to
look at a specific entry. This error occurs if the index provided does not point to
a valid entry
–224 Illegal parameter
value; Invalid list
length
You are trying to set some list measurement settings, but the multiple lists that
you sent had differing lengths. The number of settings must be consistent from
list to list
–224 Illegal parameter
value; LXI Event
<event> already
exists.
This error occurs when you try to add an LXI Event that has already been added
–224 Illegal parameter
value; LXI Event
<event> contains
illegal characters.
When a new LXI Event is created, it may not use the comma, semicolon, or
newline characters. All other printable ASCII characters are valid
–224 Illegal parameter
value; LXI Event
<event> does not
exist.
The requested event has not been added yet