100 Keysight N9010B EXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–214 Trigger deadlock The trigger source for the initiation of a measurement is set to GET, and the
following measurement query was received. The measurement cannot be
started until a GET is received, but the GET would cause an INTERRUPTED
error
–215 Arm deadlock The arm source for the initiation of a measurement is set to GET and the
following measurement query is received. The measurement cannot be started
until a GET is received and the GET would cause an INTERRUPTED error
–220 Parameter error A problem was found with a program data element. The exact problem cannot
be specifically identified
–221 Settings conflict;
There are many types of settings conflict errors. For full details, see
“–221
Settings Conflict Errors” on page 85
–222 Data out of range; A data element was found but the instrument could not be set to that value
because it was outside the range defined for the command. A descriptive
message may be appended, such as “clipped to upper limit”
–222 Data out of Range;
clipped to source
max/min
A source parameter has been entered that exceeds the range of the selected
source. The parameter has been clipped to match the range of the source
–222 Data out of range;
Invalid list data
You tried to use a trace that has a number of sweep points that differs from the
current sweep points setting
–222 Data out of range;
Two entries already
exist at this x-axis
value.
When entering values for limit lines, you cannot have more than two Y-axis
(amplitude) values entered for a specific X-axis (frequency) value
–223 Too much data A data element (of block, expression, array type, or string type) had more data
then allowed by the command, or by the available memory
–223 Too much data; 200
spurs found.
Additional spurs
ignored.
There are too many spurs for the table (the limit is 200), and any additional
spurs that are found will be ignored
–224 Illegal parameter
value
An exact data value (from a list of the allowed values) was required, but not
found. See the specific feature description for information about the expected
parameter values
–224 Illegal parameter
value; <Value>
invalid. Fractional
values are not
allowed.
The seconds parameter of an LXI time may not contain a fractional portion. For
example 123456789.0 is valid, but 123456789.1 is not valid
–224 Illegal parameter
value; <value> out of
range.
The value does not fall within the valid range
Err# Message Verbose/Correction Information