Keysight N9010B EXA Signal Analyzer Service Guide 99
Instrument Messages
Event Messages
–200 Execution error;
Trace file created by
incompatible version
of Phase Noise App
The trace file may have been created by another version of the Phase Noise
personality, which uses a different trace format that is incompatible with the
version you are running. Please check that you are running the most
up-to-date version of the personality
–200 Function not
available before
Marker Zoom is
performed
Function not available before perform Marker Zoom
–200 Must perform Scan
before do Search
Cannot perform Search as the trace data is empty
–200 No marked signal Cannot perform the selected function because no signal was marked. You must
mark the peak of interest before selecting the function
–200 No Measure At Marker
Added to Signal List
No signal peak was added into Signal List as there is no valid measure to
Marker result. You must perform Measure at Marker before selecting Measure
At Marker --> List
–200 No Peak Added to
Signal List
No signal peak was added from the Search to Signal List as no signal peaks
were found within the defined parameters of the search criteria
–200 Signal List is Empty Cannot perform the selected function because the signal list contains no data
–200 Signal List is Full Cannot perform the selected function because the signal list is full. Please
clear the list
–200 Signal Selected is
not in the Signal
List
The signal selected is not the list. You only can perform the operation on signal
that is already available in the list
–201 Invalid while in
local
The command cannot be executed while the instrument in Local control
–202 Settings lost due to
rtl
A “return to local” control was forced and some settings were lost as a result of
this
–203 Command protected The command could not be executed because it is disabled. It was disabled by
licensing or password protection
–203 Command protected;
feature not licensed
The specified feature, for example “N9073A-TR2” is not licensed. The license
may have expired. You cannot use it until you obtain a license
–210 Trigger error A trigger error has occurred, but the exact problem cannot be specifically
identified
–211 Trigger ignored A GET, *TRG or other triggering signal was received, but was ignored because
of timing considerations. For example, maybe the instrument was not ready to
respond when the command was received
–212 Arm ignored An arming signal was received, but it was ignored
–213 Init ignored An initiate trigger/sweep request was received and ignored, because another
measurement was already in progress