98 Keysight N9010B EXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–200 At Full Zoom Marker Zoom is not available as it has reached full zoom
–200 Execution Error A program execution error has occurred. The exact problem cannot be
specifically identified
–200 Execution error;
Carrier frequency
outside device’s
transmit band
The entered channel/carrier frequency is not within the range of your current
mode setup selection of standard and device
–200 Execution error;
Invalid GSM burst
timing
A GSM-like burst was acquired, but its timing is not valid. Ensure the correct
Burst Type has been selected
–200 Execution error;
Invalid IP address
The IP address supplied is either not valid or does not belong to a compatible
Signal Generator. Please check the IP address and instrument connection and
try again
–200 Execution error;
Invalid Marker Trace.
Cannot place markers on the reference trace, because the reference trace is
currently turned off or has no data
–200 Execution Error; No
peak found.
No signal peak was found within the defined parameters of the search. (Note:
for ESA/PSA compatibility, the Swept SA measurement uses 780 for this error
number; all other measurements use –220)
–200 Execution error; No
ranges are defined.
Activate a range.
There are no active ranges in the range table. You will need to activate at least
one range
–200 Execution Error;
Preselector centering
failed
Algorithm failed to center the preselector. This maybe caused by the signal
peak being too low in amplitude, or it could be due to excessive CW input
signal, alignment error, or hardware failure
–200 Execution Error;
Signal not stable
enough to track
The signal that you have selected to track is changing too much for the function
to track it properly
–200 Execution Error;
Store ref trace
before turning on
Normalize
The Reference trace data must be stored in the Ref trace before you turn on the
Normalization function
–200 Execution error; Sync
word was not found.
NADC & PDC: In an EVM measurement, the sync word is not found and the
synchronization cannot be established when Sync Word is selected in the Burst
Sync menu.
Flexible Digital Demodulation: The sync word cannot be detected because of
inappropriate parameter settings or incorrect signal
–200 Execution error;
Trace file contains
no compatible traces.
The trace file may have been created by another version of the Phase Noise
personality, which uses a different trace format that is incompatible with the
version you are running. Please check that you are running the most
up-to-date version of the personality
Err# Message Verbose/Correction Information