140 Keysight N9010B EXA Signal Analyzer Service Guide
RF Section Troubleshooting (RF/Microwave Analyzers)
RF Section Description
A13 RF Front End Assembly
This assembly is a self-contained microcircuit that is repaired at the assembly
level. See Figure 4-1.
Figure 4-1 A13 RF Front End Assembly View from Front Panel
This assembly contains the following circuits:
Input Low-pass filter (RF input signals < 3.6 GHz)
—Low Band Limiter
Optional Electronic Attenuator (0-24 dB attenuation control)
Optional Low Band Preamplifier
RF Mixer #1 (RF input signals < 3.6 GHz)
1st L.O. Sub-system (3.8 GHz to 8.7 GHz)
1st I.F. Amplifier (5122.5 MHz)
I.F. Band pass filter (5122.5 MHz)
I.F. Low-pass filter (5122.5 MHz)
Mixer #2 (RF input signals 3.6 GHz to 13.599 GHz)
Microwave Input Amplifier
Mixer #3 (RF input signals 13.6 GHz to 26.5 GHz)
2nd I.F. Amplifier (322.5 MHz)
Microwave L.O. Sub-system (3.8 GHz to 8.7 GHz)
L.O. Doubler
(Option EXM only)