164 Keysight N9010B EXA Signal Analyzer Service Guide
RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
Input Attenuator B Power Level Verification
Press AMPTD, Attenuation, Mech Atten, 0 dB. Remove output cable W9 from
A10 (2). Refer to Figure 4-13. Measure the 50 MHz calibrator signal on the
output of the attenuator using a functioning Spectrum Analyzer.
The level should be 25 dBm ± 2 dB as shown in Figure 4-15.
Figure 4-15 50 MHz Calibrator Signal on Output of Attenuator B