Keysight N9010B EXA Signal Analyzer Service Guide 191
RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
The level should be 28 dBm ± 2 dB as shown in Figure 4-32.
Figure 4-32 4.8 GHz Calibrator Signal on Output of Attenuator B
If the analyzer is equipped with Option FSA, press Mech Atten and enter 6 dB.
The 4.8 GHz calibrator signal measured on the functioning Spectrum Analyzer
should measure 6 dB lower than the previous step (~34 dBm).
Press Mech Atten and enter 10 dB. The 4.8 GHz calibrator signal measured on
the functioning Spectrum Analyzer should measure 10 dB lower than the first
step (~38 dBm). Continue testing all the attenuation settings as described in
Table 4-18.
Table 4-18
Attenuator Setting Expected Level
dBm @ 4.8 GHz
Tolerance
16
a
44 ±3 dB
20 48 ±3 dB
30 58 ±3 dB
40 68 ±4 dB
50 78 ±4 dB
60 88 ±4 dB
a. requires Option FSA