212 Keysight N9010B EXA Signal Analyzer Service Guide
RF Section Troubleshooting (Millimeter-Wave Analyzers)
Troubleshooting
A9 Input Attenuator and A10 Input Attenuator Verification
Calibrator Switch Test
On the A16 Reference assembly, disconnect semi rigid cable W3 from A16J701
and measure A16J701 with a spectrum analyzer. Expected signal is 50 MHz at
25 dBm ± 0.5 dB (RF calibrator set to 50 MHz). If signal level is incorrect,
suspect A16 Reference assembly is faulty, or the 50 MHz calibrator amplitude
requires adjustment using the field calibration software.
To verify calibrator switch operation, connect external signal source set to
50 MHz and 25 dBm to the RF input connector of analyzer under test. Press
Input/Output, RF Calibrator, Off. If the signal level at the attenuator output is
now correct, suspect the A9 Attenuator cal switch or a faulty control signal
from the A15 Front End Controller assembly.
Attenuator Check
Set the signal source connected to the analyzer input port to 0 dBm. On the
analyzer, press Amplitude, Attenuation, and change input attenuation to 0 dB.
The measuring spectrum analyzer connected to A10 Attenuator output port
should indicate a 0 dBm level.
Change the input attenuation on the analyzer under test to 2 dB (if Option FSA
is present) or to 10 dB (if Option FSA is not present). See the chart below for
expected measurement values at the A10 Attenuator output port.
Analyzer Under Test
Attenuator Setting (dB)
Power at A10
Output Port (dBm)
Input Attenuator Being Tested
(Options 532, 544)
0 0 dBm (reference) Both set to through path
2
a
a. These settings only available if Option FSA is installed.
2A9
4
a
4A9
6
a
−6 A9
8
a
8A9
10 10 A10
20 20 A10
30 30 A10
40 40 A10
50 50 A10
60 60 A10