Keysight N9010B EXA Signal Analyzer Service Guide 459
Assembly Replacement Procedures
RF Area (Option 503, 507, 513, 526)
3. Refer to Figure 15-14. Remove semi-rigid coax cable W35 and the fixed
attenuator (1).
Figure 15-14 W35 and Attenuator Removal
4. Remove ribbon cable W18 from J4 of the A11 Low Band Switch.