462 Keysight N9010B EXA Signal Analyzer Service Guide
Assembly Replacement Procedures
RF Area (Option 503, 507, 513, 526)
7. Refer to Figure 15-17. Using the T-8 driver, remove the two 0515-1992
screws (3) to separate the #1 switch/bracket assembly (1)/(2) from the
low band switch bracket (4).
Figure 15-17 Switch 1 Removal