Keysight N9010B EXA Signal Analyzer Service Guide 487
Assembly Replacement Procedures
RF Area (Option 532, 544)
Replacement
1. Refer to Figure 15-41. Place the switch onto the bracket and replace the
three screws (1). Torque to 6 inch-pounds.
2. Refer to Figure 15-40. Place the switch/bracket into place onto the
chassis and replace the two screws. Torque to 9 inch-pounds.
3. Plug in the ribbon cable into the A15 Front End Control assembly.
4. Refer to Figure 15-39. Install W3, W45, W46, W47, and W48. Torque to
10 inch-pounds.