Keysight N9010B EXA Signal Analyzer Service Guide 89
Instrument Messages
Event Messages
–221 Settings conflict;
Freq > 3.6 GHz
unavailable while
electronic attenuator
enabled
The electronic attenuator does not function above 3.6 GHz, so if you have
that attenuator enabled, you cannot change the center frequency so that
frequencies above 3.6 GHz are displayed/measured
–221 Settings conflict;
Function not
available in External
Mixing
The frequency offset feature cannot be used when you have selected a log
scale for the frequency axis
–221 Settings conflict;
Function not
available in Tracking
Source mode
The feature cannot be used when you are using a Tracking Source
–221 Settings conflict;
Function only
available in Tracking
Source mode
The feature cannot be used unless you are using a Tracking Source
–221 Settings conflict;
Function unavailable
with MW Presel off
You cannot center or adjust the preselector because the Microwave
Preselector is currently off
–221 Settings conflict;
Gate control is Edge
for Gated FFT
You cannot use level triggering to control the gate if you are using the FFT
gating method
–221 Settings conflict;
Gate control must be
Edge for this Gate
Source
You cannot use level triggering to control the gate when you are using the
currently selected gate source
–221 Settings conflict;
Gate is not available
when Marker Count on
The gate function cannot be used while you have marker count turned on
–221 Settings conflict;
Gate Length is not
settable in FFT
sweeps
The sweep time for FFT sweeps is set by the calculations, so sweep time
settings cannot be adjusted
–221 Settings conflict;
Gate Length is not
settable in FFT
sweeps
The sweep time for FFT sweeps is set by the calculations, so sweep time
settings cannot be adjusted
–221 Settings conflict;
Gate Method is not
compatible with
current Sweep Type
setting
If the Gate is On and you have the FFT Sweep Type manually selected, then
the Gate Method cannot be selected
Err# Message Verbose/Correction Information