90 Keysight N9010B EXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–221 Settings conflict;
Gate not available
with external
Tracking Source
The Gate functions are unavailable when Source Mode is Tracking with an
external source. This is because the Gate circuitry is used to sync the external
source
–221 Settings conflict;
Gate not available
with Tracking
Generator
If the Source Type is Tracking Generator, the Gate circuitry is used for TG
sync and is not available for gating
–221 Settings conflict;
Gated FFT is not
available while Sweep
Type is set to Swept
The gated FFT function is not available if you have selected the swept type of
sweep. You must be in the FFT sweep type
–221 Settings conflict;
Gated LO is not
available while Sweep
Type is set to FFT
The FFT sweep type moves the LO frequency in steps. So the gated LO
function is not available if you have selected FFT sweep
–221 Settings conflict;
Gated Video is not
available while Sweep
Type is set to FFT
The gated video function is not available if you have selected the FFT sweep
type
–221 Settings conflict;
Incorrect RBW for
demod. Change RBW
–221 Settings conflict;
Ind I/Q is not
available for this
measurement
The Independent I and Q setting is not available for the current
measurement. Only some measurements (initially, only VXA) support this
setting
–221 Settings conflict;
Invalid trace number
The subopcode used to specify the trace number is invalid for this
measurement or query
–221 Settings conflict;
Knob is not available
to modify this
function
You should select a specific value for this function. Scrolling through values
with the knob is not allowed
–221 Settings conflict; LO
Phase Noise Adj not
available
For instruments without the Dual-Loop LO, this feature is not available
–221 Settings conflict;
Marker 1 Trace
Update=off turns off
Signal Track
Signal Track not available unless the trace containing Marker 1 is updating
–221 Settings conflict;
Marker cannot be
relative to itself
A marker must be set relative to another marker, not to itself
Err# Message Verbose/Correction Information