Keysight N9010B EXA Signal Analyzer Service Guide 91
Instrument Messages
Event Messages
–221 Settings conflict;
Marker Count is not
available when Gate
on
The marker count function cannot be used while you have gating turned on
–221 Settings conflict;
Marker Function is
not available for a
Fixed marker
If a Marker is a Fixed type marker, the marker's value does not change from
when it first became fixed. You cannot turn on or change a Marker Function
because there is no ongoing measurement data to use for the marker
function calculation
–221 Settings conflict;
Marker type must be
delta
Mkr?->Span and Mkr?->CF require that the selected marker be a delta
marker
–221 Settings conflict;
Marker-> function is
not available in zero
span
Most of the
Marker To functions are not available if you are in zero span
(span = 0 Hz, or time domain), so you cannot send the commands for these
functions
–221 Settings conflict;
Mask unavailable for
current Span.
Increase to display
mask.
The current span setting is either narrower than the mask width or so wide
that there are too few display points to allow the mask to be drawn. Increase
or decrease the span to display the mask
–221 Settings conflict;
Meas Type was changed
to Examine for Exp
Avg Mode.
Average Mode has been changed to Exponential. Full Meas Type is not
available for Exponential Average Mode, therefore Meas Type has been
changed to Examine
–221 Settings conflict;
Meas Type was changed
to Full for Repeat
Avg Mode
Average Mode has been changed to Repeat. Examine Meas Type is not
available for Repeat Average Mode therefore Meas Type has been changed to
Full
–221 Settings Conflict;
MinPts/RBW limit not
met
–221 Settings conflict;
Mkr -> CF is not
available when the
x-axis is time domain
The marker to center frequency functionality does not work when the X-axis
is in the time domain
–221 Settings conflict;
Mod Scheme AutoDet
unavailable when
Burst Sync=RF Amptd
The modulation scheme auto detection is unavailable when the burst sync is
set to RF Amptd (GSM/EDGE)
–221 Settings conflict; MS
gain is not available
in this Mode
Mobile Station gain correction is not available in some Modes, or in some
measurements (for example, the SA measurement)