Keysight N9010B EXA Signal Analyzer Service Guide   91
Instrument Messages
Event Messages
–221 Settings conflict; 
Marker Count is not 
available when Gate 
on
The marker count function cannot be used while you have gating turned on
–221 Settings conflict; 
Marker Function is 
not available for a 
Fixed marker
If a Marker is a Fixed type marker, the marker's value does not change from 
when it first became fixed. You cannot turn on or change a Marker Function 
because there is no ongoing measurement data to use for the marker 
function calculation
–221 Settings conflict; 
Marker type must be 
delta
Mkr?->Span and Mkr?->CF require that the selected marker be a delta 
marker
–221 Settings conflict; 
Marker-> function is 
not available in zero 
span
Most of the 
Marker To functions are not available if you are in zero span 
(span = 0 Hz, or time domain), so you cannot send the commands for these 
functions
–221 Settings conflict; 
Mask unavailable for 
current Span. 
Increase to display 
mask.
The current span setting is either narrower than the mask width or so wide 
that there are too few display points to allow the mask to be drawn. Increase 
or decrease the span to display the mask
–221 Settings conflict; 
Meas Type was changed 
to Examine for Exp 
Avg Mode.
Average Mode has been changed to Exponential. Full Meas Type is not 
available for Exponential Average Mode, therefore Meas Type has been 
changed to Examine
–221 Settings conflict; 
Meas Type was changed 
to Full for Repeat 
Avg Mode
Average Mode has been changed to Repeat. Examine Meas Type is not 
available for Repeat Average Mode therefore Meas Type has been changed to 
Full
–221 Settings Conflict; 
MinPts/RBW limit not 
met
–221 Settings conflict; 
Mkr -> CF is not 
available when the 
x-axis is time domain
The marker to center frequency functionality does not work when the X-axis 
is in the time domain
–221 Settings conflict; 
Mod Scheme AutoDet 
unavailable when 
Burst Sync=RF Amptd
The modulation scheme auto detection is unavailable when the burst sync is 
set to RF Amptd (GSM/EDGE)
–221 Settings conflict; MS 
gain is not available 
in this Mode
Mobile Station gain correction is not available in some Modes, or in some 
measurements (for example, the SA measurement)
Err# Message Verbose/Correction Information