RF Section Troubleshooting
Troubleshooting
Input Attenuator A Power Level Verification
Press Mode Preset, Input/Output, RF Calibrator, 50 MHz, AMPTD,
Attenuation, 0 dB on the analyzer. Refer to Figure 4-13, remove cable W11
from A9 (1) Output. Measure the 50 MHz calibrator signal on the output of the
attenuator using a functioning Spectrum Analyzer.
Figure 4-13 Cable W11