172 
RF Section Troubleshooting
Troubleshooting
Figure 4-17  Cable W3 Location
When the analyzer is tuned to a center frequency of 50 MHz, the Low Band 
switch should have minimal loss. Press Input/Output, RF Calibrator, 50 MHz, 
AMPTD, Attenuation, 10 dB on the analyzer. Measure the 50 MHz calibrator 
signal on the output of A11J2 using a functioning Spectrum Analyzer. The level 
should be −35 dBm ± 5 dB as shown in Figure 4-18.