171
RF Section Troubleshooting
Troubleshooting
NOTE 1. It may be difficult to measure the higher attenuator settings using the −25 
dBm internal calibrator signal. Use an external source with the frequency set 
to 50 MHz and adjust the output level to 0 dBm. This will increase the 
measured power levels noted in the table above by 25 dB.
2. Tolerances should be used as a guideline. 
If the power levels measure correctly, reconnect the W9 cable and reset the 
input attenuation to 10 dB by pressing Mech Atten, 10 dB on the analyzer.
If either of these levels measure incorrectly, Input Attenuator B is the most 
probable cause, provided the switch control logic has been verified. 
Low Band Switch Power Level Verification 
Refer to Figure 4-17. Disconnect the W3 cable at A11J2 of the A11 Low Band 
Switch (1).