Analog/Digital IF Troubleshooting
25 MHz BW IF Section
NOTE If the analyzer is not in 10 dB of input attenuation press AMPTD, Attenuation, 
10 dB. 
9. Refer to Figure 6-9, carefully disconnect W12 at A3J14.
Figure 6-9  A3 Digital I.F. Cables
10.Connect the W12 cable to the MMCX female to SMA female connector. 
Use an appropriate cable to go from the SMA connector to the RF input of 
a functioning spectrum analyzer.
11.Press Freq, 10 MHz, Span, 1 MHz, Peak Search on the functioning 
spectrum analyzer.