257
Analog/Digital IF Troubleshooting
40 MHz BW IF Section
11.The analyzer should read 322.5 MHz at −23 dBm ± 3 dB as shown in Figure 
6-12.
Figure 6-12  A15 322.5 MHz Output
NOTE If the 322.5 MHz signal is not measuring the correct power level, refer to 
Chapter 4, “RF Section Troubleshooting” in this service guide.
Reconnect W37 to A15J900.
IMPORTANT Measuring at this location is for convenience. The 322.5 MHz signal is an 
output at A15J900. Since the W37 cable connects from this output to the 
input of the Analog I.F. assembly at A2J100, the small coaxial cable has not 
been tested at the A2J100 end.