267
Analog/Digital IF Troubleshooting
40 MHz BW IF Section
12.The analyzer should read 22.5 MHz at 25 dBm ± 4 dB as shown in Figure
6-18.
Figure 6-18 22.5 MHz IF Input
13.If the 22.5 MHz signal is not measuring the correct power level, see “A2
Analog I.F. Troubleshooting” on page 236 in this service guide.
14.If the 22.5 MHz signal is within tolerance, carefully reconnect the W41
cable to A3J19.