568
Post-Repair Procedures
Post-Repair Procedures
S/W Revision Greater Than or Equal To A.01.55 and Less Than A.02.00
1. Close the instrument application software by pressing
File, Exit, Enter
using the front panel keys.
2. Connect a USB mouse to one of the instrument front panel USB ports.
3. Connect a USB FLASH drive to the other front panel USB port.
4. Using the mouse double-click on the Windows Explorer link on the
desktop and navigate to the following folder:
C:\Program Files\Agilent\SignalAnalysis\Physics
5. Double-click on BackupAndRestore.exe and the window shown in
Figure 15-1 will appear.
Figure 15-1 Calibration Data Backup
6. Under the “Specify Backup Folder or File Name” use the mouse to
select the “...” button to open the “Browse For Folder” dialog box.
7. Select the USB FLASH drive and press
OK.
NOTE While you can use the “Make New Folder” button in the “Browse For
Folder” dialog box to create a sub-folder on the USB FLASH drive, and you
can specify a filename in the “Specify Backup Folder or File Name”
window, you can also just use the root directory on the drive and let the
instrument select a default filename for you.
8. Select “Backup” and wait for the calibration database to be backed up.
9. When the backup has completed close the Calibration Data Backup
And Restore window.