614
Functional Tests
Amplitude Accuracy at 50 MHz
Input/Output, RF Input, RF Coupling, select DC
Sweep/Control, Sweep Setup, Swp Time Rules, SA - Accuracy
Meas Setup, Average/Hold Number, 20, Enter
Trace/Detector, Trace Average
Single
7. Perform the following steps for each row listed in Table 16-5:
a. Set the synthesized sweeper amplitude to the value listed in the
Nominal Source Amplitude column in Table 16-5
b. Set the Mech Atten as indicated in the Attenuation column in Table
16-5
c. Set the Span as listed in the Span column of Table 16-5
d. Record the source amplitude, as measured by the power meter, in
the Power Meter Amplitude column of Table 16-5
e. On the signal analyzer, press Restart.
f. Wait for the signal analyzer to finish averaging.
g. Press Peak Search.
h. Record the signal amplitude, as measured by the analyzer in the
Measured Amplitude column of Table 16-5
i. Calculate the signal amplitude accuracy error using the following
equation, and record the results under the Amplitude Accuracy
Error column:
Amplitude Accuracy Error = Meas_amp - Power_meter
Table 16-5 Amplitude Accuracy Results
Nominal
Source
Amplitude
(dBm)
Attenuation
(dB)
Span
(MHz)
Measured
Amplitude
Meas_amp
(dBm)
Power Meter
Amplitude
Power_meter
(dBm)
Amplitude
Accuracy
Error
(dB)
Test Limit
(dB)
-4102 ±1.13dB
-9101 ±1.13dB
-14100.5 ±1.13dB
-4200.1 ±1.13dB
-14200.1 ±1.13dB
-4300.1 ±1.13dB
-14 30 0.1 ±1.13 dB