EasyManua.ls Logo

Keysight N9030B - Burst Width Auto Detection; Burst Width

Keysight N9030B
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Spectrum Analyzer Mode
3.7 Burst Power Measurement
Remote Command
[:SENSe]:TXPower:METHod THReshold | BWIDth
[:SENSe]:TXPower:METHod?
Example
:TXP:METH BWID
:TXP:METH?
Dependencies Only appears in Spectrum Analyzer and GSM/EDGE Modes
Preset
THReshold
State Saved Yes
Range Above Threshold Lvl|Burst Width
Backwards Compatibility
SCPI
[:SENSe]:BPOWer:METHod
Burst Width Auto Detection
Turns Burst Width Auto Detection on or off.
Remote Command
[:SENSe]:TXPower:BURSt:AUTO ON | OFF | 1 | 0
[:SENSe]:TXPower:BURSt:AUTO?
Example
:TXP:BURS:AUTO 1
:TXP:BURS:AUTO?
Preset
ON
State Saved Yes
Backwards Compatibility
SCPI
[:SENSe]:BPOWer:BURSt:AUTO
Burst Width
When "Burst Width Auto Detection" on page 1324 is OFF, you may enter a fixed-
time value in seconds, or alternatively specify the burst width as a percentage of the
last measured burst width (result in bottom-left corner of second window).
Remote Command
[:SENSe]:TXPower:BURSt:WIDTh <time>
[:SENSe]:TXPower:BURSt:WIDTh?
Example
:TXP:BURS:WIDT 10
:TXP:BURS:WIDT?
:TXP:BURS:AUTO 0
:TXP:BURS:AUTO?
Dependencies Grayed-out if "Meas Method" on page 1323 is ‘Above Threshold Lvl’
Gray-out message: ‘Settings Conflict; Meas Method must be Measured Burst Width’