3 Spectrum Analyzer Mode
3.7 Burst Power Measurement
3.7.9.1 Sweep/Control
Accesses controls that let you operate the sweep and control functions of the
instrument, such as Sweep Time and Continuous/Single.
Sweep Time
Sets the sweep time to capture and show on screen.
Remote Command
[:SENSe]:TXPower:SWEep:TIME <time>
[:SENSe]:TXPower:SWEep:TIME?
Example
:TXP:SWE:TIME 10
:TXP:SWE:TIME?
Dependencies Only appears in Spectrum Analyzer Mode
Preset 640 us
State Saved Yes
Min 1.0e-6
Max 50
Sweep/Measure
Lets you toggle between Continuous and Single sweep or measurement operation.
The single/continuous state is Meas Global, so the setting affects all measurements.
The front-panel key Single/Cont performs exactly the same function
See "More Information" on page 1344
Remote
Command
:INITiate:CONTinuous OFF | ON | 0 | 1
:INITiate:CONTinuous?
Example Put instrument into Single measurement operation:
:INIT:CONT 0
:INIT:CONT OFF
Put instrument into Continuous measurement operation:
:INIT:CONT 1
:INIT:CONT ON
Preset
ON
Note that :SYST:PRES sets :INIT:CONT to ON, but *RST sets :INIT:CONT to OFF
State Saved Saved in instrument state
Annunciation The Single/Continuous icon in the Meas Bar changes depending on the setting:
Spectrum Analyzer Mode User's &Programmer's Reference 1343