EasyManua.ls Logo

Keysight N9030B - 6.3.5 Q Ext Gain; 6.3.6 Q Gain in I+jQ

Keysight N9030B
2852 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Input/Output
6.3 External Gain
6.3.5 Q Ext Gain
Affects the Q channel input.
Remote Command
[:SENSe]:CORRection:IQ:Q:GAIN <rel_ampl>
[:SENSe]:CORRection:IQ:Q:GAIN?
Example Set the Q Ext Gain to 10 dB:
:CORR:IQ:Q:GAIN 10
Set the Q Ext Gain to -10 dB (that is, a loss of 10 dB):
:CORR:IQ:Q:GAIN -10
Dependencies Not available unless option BBA is installed
Grayed-out when Q gain in I+jQ is set to Same as I Gain
Preset 0 dB
Unaffected by Preset, but set to 0dB by Restore Input/Output Defaults or Restore System Defaults-
>All
State Saved Saved in instrument state
Min/Max –/+100dB
Annotation Ext Gain: <Q Ext Gain> dB
No annotation is shown when Input is not I/Q. Also not shown when Q Ext Gain is 0.00 dB. Q Ext Gain
is not shown for Input Path I Only or I+jQ. When Input Path is Independent I and Q and when I and Q
Ext Gain are both non-zero but are the same the annotation will be "Ext Gain: <Ext Gain> dB" and
when I Ext Gain is not the same as Q Ext Gain, both are shown. "Ext Gain: <I Ext Gain> dB, <Q Ext
Gain> dB"
6.3.6 Q Gain in I+jQ
When Same as I Gain (ON) is selected, I Ext Gain value is applied to both I and Q
channel input if the Input Path is I+jQ.
When Independent (OFF) is selected, I and Q Ext Gain values are applied to I and Q
channel input independently.
Remote
Command
[:SENSe]:CORRection:IQ:Q:GAIN:COUPle ON | OFF | 0 | 1
[:SENSe]:CORRection:IQ:Q:GAIN:COUPle?
Example
:CORR:IQ:Q:GAIN:COUP ON
:CORR:IQ:Q:GAIN:COUP?
Preset
ON
State Saved Yes
Range Same as I Gain | Independent
Spectrum Analyzer Mode User's &Programmer's Reference 2405