3 Spectrum Analyzer Mode
3.2 Swept SA Measurement
Remote Command
:CALCulate:LIMit[1]|2|…|6:CLEar
Example
:CALC:LIM2:CLE
deletes all data for limit line 2
Couplings Identical to :CALC:LLIN:DEL
Trace Fail (Remote Query Only)
Tests a trace against all associated limit lines. Returns 0 if the trace is within all
limits and margins, 1 if the trace exceeds either the limit or the margin. If no limits
apply to the selected trace, automatically returns 0.
Only applies to limits that are turned on, if a Limit is off it will not be tested. If a Trace
is not displaying, it will still be tested, and if Test Limits (All Limits) is off the Trace
will still be tested.
This command ignores limit lines that are assigned to other traces.
Remote
Command
:CALCulate:TRACe[1]|2|…|6:FAIL?
Example
:CALC:TRAC3:FAIL?
returns zero if there is no failure, 1 if the trace exceeds either the limit or the margin
3.2.8.3 Meas Standard
Includes controls for setting the Radio Standard for PowerSuite measurements, as
well as controls for setting the EMC Standard and CISPR Preset to which the current
measurement will be made.
The EMC Standard and CISPR Preset controls only appear with Option EMC or the
N6141A application or the W6141A application installed and licensed.
Dependencies Only appears if you have the proper option installed in your instrument
Radio Standard Presets
Lets you specify the Radio Standard to be used. Spectrum Analyzer Mode supports
many Radio Standards. You can select the desired Radio Standard using the Radio
Std Presets dialog, which is a cascading list of standards. When you have the
selected the desired Standard, press OK and the measurement settings will change
to reflect that standard.
Remote Command
[:SENSe]:RADio:STANdard[:SELect] NONE | JSTD | IS95a | IS97D | IS98D | GSM |
W3GPP | C2000MC1 | C20001X | NADC | PDC | BLUEtooth | TETRa | WL802DOT11A |
WL802DOT11B | WL802DOT11G | HIPERLAN2 | DVBTLSN | DVBTGPN | DVBTIPN | FCC15